Waters 2545 Binary Gradient Module Operator's Manual page 17

Table of Contents

Advertisement

Key (Continued)
Number
Name
3
Filter
4
Seal-wash B outlet tube
5
Drip tray
6
Drain tube
7
Inlet manifold B
8
Seal-wash B connection tube
9
Seal-wash B inlet tube
10
Vent valve
11
Seal-wash pump B
12
Waste outlet at tee fitting
13
Seal-wash A outlet tube
14
Seal-wash A connection tube
15
Inlet manifold A
If you plan to run only two solvents, make sure to plumb one solvent into
each manifold and connect the tube to a port in each manifold.
You can shorten the solvent inlet tubes to accommodate the solvent
container locations.
5.
Insert the filter ends of the solvent tubes into the appropriate solvent
container.
Caution:
!
solvent containers for reliable operation. If using pressurized
containers for blanketing large containers of solvent, do not exceed
5 psig. Pressures above 5 psi can force solvent through the internal
components of the solvent selection valve (SSV), possibly causing a
spillover into the solvent waste reservoir or an adverse mixture
with other solvents.
The binary gradient module does not require pressurized
Making plumbing connections
1-11