Unpacking and inspecting

The binary gradient module is shipped in a single carton on a wooden pallet.
Save the carton and pallet in case you transport or ship the unit later.
To unpack and inspect the binary gradient module:
1.
Ensure that there is sufficient bench space for the binary gradient
module.
2.
Check the outside of the shipping carton for signs of damage. Document
or photograph areas of damage.
3.
Remove the plastic wrap, if any, and the bands securing the carton to
the pallet. Remove the top and the sides of the shipping carton, the
Startup Kit, and the packing material.
4.
Check that the contents of the Startup Kit match the Startup Kit list.
Caution:
!
the unit.
Warning:
module from the pallet to the bench. Most of the module's weight is
at the front.
5.
Remove the binary gradient module from the bottom styrofoam cushion
and the box bottom, and place it on a level, clean surface, making sure to
leave room for the other system components.
6.
Check that the instrument serial number (found on the rear panel near
the power cord connector) corresponds to the number on the Certificate
of Validation.
7.
If you detect any damage to the contents of the carton, refer to
Shipments, Damages, Claims, and Returns, in Waters Licenses,
Warranties, and Support Services.
8.
Measure the voltage at the receptacle, and verify that the voltage is
between 90 and 264 Vac.
1-6
Setting up the 2545 Binary Gradient Module
To prevent damage, lift the module by the bottom sides of
To avoid injury, ensure that at least two people lift the