Adaptive Interference Suppression - HBM PW15iA Mounting Instructions

Digital load cell
Table of Contents

Advertisement

Available languages

Available languages

Output format
Binary, 2 chars. (INT)
Binary, 4 chars. (LONG)
ASCII
PW15iA
S activate an automatic zero on start up function,
S activate an automatic zero tracking function,
S linearize the scale curve with a third order polynomial,
S activate various digital filters. Available filters include
those with cut-off frequencies below 1 Hz, fast-settling
filters for dynamic measurements, notch filters and
mean value filters.
Use MSV? to read out the current measured value. The
format of the measured value (ASCII or binary) is set
with the COF command. You can also use the COF com­
mand to activate automatic measurement output. The
measured values are transmitted in the following format,
subject to the COF command:
Input signal
0 ... maximum
capacity
0 ... maximum
capacity
0 ... maximum
capacity
3.3

Adaptive interference suppression

Whatever the mode of operation, you can use the ADF
command to activate automatic interference suppression
with adaptive filters. Interference frequencies are auto­
matically found during measurement and suppressed by
comb filters and averaging. The maximum filter settling
time can be limited with the TMA command.
A4359-1.0 HBM: public
Structure and mode of operation
Output when NOV=0
0 ... 20,000 digits
0 ... 5,120,000 digits
0 ... 1,000,000 digits
Output when
NOV>0
1 ... NOV
1 ... NOV
1 ... NOV
13

Advertisement

Table of Contents
loading

Table of Contents