Mounting
Secure the load cell in the mounting holes. A platform for
load application can be mounted on top. The screws and
tightening torques to be used are given in the table
below:
Thread
Min. property
class
M6
10.9
M6
10.9
Important
The unevenness of the surface at the connecting faces
must be no worse than 0.1 mm.
A4359-1.0 HBM: public
Loading direction
Mounting plate
(not included in the scope of supply)
Tightening
torque
10 N @ m
10 N @ m
Load application
(customer construction)
Max. thread reach
10 mm
10 mm