Str Act
Press "SET" to access the parameter setting.
Str=Auto
To change the setting to always on press the "UP" or "DOWN" buttons and
then "SET". When set to "Str=On" the stirrer motor comes on with the bath
power regardless of the temperature set in the stirrer set-point parameter.
8.11.3

Stir set-point

Allows setting of the temperature above which the stirrer motor will activate
when the stirrer activation is set to automatic.
To access the paramter press "SET" from the stirrer activation parameter.
Set-point mode selection parameter is indicated by,
Str SEt
Press "SET" to access the parameter value.
Str=200
Press "UP" or "DOWN" to change the value and then "SET" to enter the new
value.
NOTE: Stir Mode Select defaults to "
bath is cycled off and back on. Therefore, this parameter has to be set
each time the bath is powered on if "
eration.
8.11.4

Heat up power

NOTE: This section applies to controller parameters, not to the High/Low
setting of the front panel heater switch.
This feature helps lengthen the lifetime of the heaters when using heat transfer
salt as the bath fluid. The heater power will be limited to less than 30% while
the salt is solid at low temperatures to prevent the heaters from overheating.
Once the temperature reaches 200°C the heaters can then operate at 100%
power.
HEAt UP
Press "SET" to access the parameter value.
Stirrer mode selection parameter
Stirrer is set for automatic activation at the stirrer set-point
temperature.
Stirrer motor activation set-point parameter
Stirrer motor activation set-point
Heat up power option
8 Controller operation
Operating parameters
" each time the power of the
" is the desired mode of op-