Temperature Set-point

Set the temperature set-point to any value within the range and resolution as given in the
specifications. Do not exceed the safe upper temperature limit of any device inserted into the well.
To set the temperature, select the set-point memory and adjust the set-point value.

Programmable Set-points

The controller stores eight set-point temperatures in memory. Quickly recall the set-points to
conveniently set the calibrator to a previously-programmed temperature set-point.
To set the temperature, select the set-point memory first. Push SET to access this function from the
temperature display function. The number of the current set-point memory shows at the left on the
display followed by the current set-point value.
Well temperature in degrees Celsius
100.0 C
Push SET
Access set-point memory
Set-point memory is currently set to 1 101 °C
1 100.
To change the set-point memory push UP or DOWN.
New set-point memory is 4 300 °C
4 300.
Push SET to accept the new selection and access the set-point value.

Set-point Value

To adjust the set-point after you select the set-point memory, push SET.
Set-point 4 value in °C
4 200.
If the set-point value is correct, push EXIT to show the well temperature. Push SET and then push UP
or DOWN to adjust the set-point value.
New set-point value
220.00
When the controller reaches the desired set-point value, push SET to accept the new value and to
access the temperature scale units selection. If you push EXIT instead of SET, any changes made to
the set-point are ignored.
Push SET
Accept new set-point value
8