Remove the plug from the mains before emergency release of the mattress support.
1) Hold the backrest.
3) Lower the backrest motor.
A minimum of 2 persons are required to emergency release a mattress section. Both persons hold the mattress
section. One person pulls out the pipe pin. Both slowly lower the mattress section until it is completely down.

4. Accessories

When 3/4 length using side rails you must always make sure that the distance between the bed end and the side rail's
handle upper edge is between 3 - < 6 cm for the the head section and more than 32 cm for the foot section
There is a risk of entrapment of fingers on assembly
and operation of the side rail.
4.1 Mounting / dismounting the Scala 2 series and Verso 11 steel side
rail
Mounting
The side rail must be mounted according to the assembly instruction side rail.
Tighten the side rail with the finger screws.
Dismounting
Loosen the finger screws and remove the side rail.
2) Remove the pipe pin from the backrest motor.
4) Lower the backrest.
1) 3 - < 6 cm
15
2) 32 cm