6.3 Maintenance

Personnel who have received the necessary instructions or training may only perform Service and maintenance of the
bed.
When relocating the bed, service must be carried out according to the service schedule.
- Before use
Ensure that all manual and electrical parts functions correctly and are in a secure state.
- After 3 months
Ensure that all manual and electrical parts are functioning, and tighten bolts, screws, nuts, etc.
- Every two year
Service should be performed after the maintenance chart. See chapter 6.2 - Maintenance chart.
Modification of Medley Ergo is strictly forbidden without authorization of Invacare

6.4 Lubrication plan

We recommend lubricating the bed according to the following instructions:
1.
Points of rotation in mattress support and base frame - Lubricate with oil.
2.
Motor attachment points to mattress platform - Lubricate with oil
3.
Upper attachment point of Rastofix fitting - Lubricate with oil
NB! The full length side rails gliding system must not be lubricated with oil – otherwise the side rail bars will move sluggishly.
Lubricate with medically clean oil, e.g. KEN-WO 50, order no.: 813239.
NB! Remove the mains plug from the socket before cleaning.