Troubleshooting - ZOLL R Series Manual

Non-invasive blood pressure
Hide thumbs Also See for R Series:
Table of Contents

Advertisement

R S
— N
-I
ERIES
ON
NVASIVE

Troubleshooting

The following table lists the error messages and problems associated with the NIBP option and
the associated corrective action(s). Read this section carefully. For further assistance, call
ZOLL Technical Service at 1-800-348-9011. International customers should call the nearest
authorized ZOLL Medical Corporation representative.
Message/Symptom
--- (dashed lines in
NIBP field)
CALIBRATE NIBP
NIBP COMM ERROR
NIBP FAULT
NIBP MEAS
ABORTED
NIBP NOT READY
NIBP OUT OF RANGE
REPEAT NIBP MEAS
26
B
P
LOOD
RESSURE
Problem
No NIBP measurement has been taken.
The last measurement taken has expired
after 120 minutes.
The patient type has been changed.
NIBP calibration is incomplete or failed
There is a communication problem with the
NIBP module.
No communication from the NIBP module.
NIBP module failed self-test.
Cuff inflation pressure is set too high for
attached cuff.
Inflation is too fast.
R Series is unable to find systolic value for
180 seconds.
Defibrillator is charged or charging. User
initiated abort.
The defibrillator is charged or charging in
progress.
NIBP module is performing power-up
self-test.
The data from the NIBP module is out of
range.
The unit exceeded the maximum number of
inflation attempts.
The unit exceeded the 180-second
measurement time limit.
www.zoll.com
(NIBP) O
I
PTION
NSERT
User Action
None, normal operation.
Repeat NIBP measurement.
Recalibrate NIBP. (See the R Series
Service Manual for details).
Cycle power and retry.
If problem persists, return for service.
Cycle power, and retry.
If the problem persists, return for
service.
Verify that you are using proper size
cuff.
Check for cuff and hose blockages.
Confirm that the unit was not charging.
If the problem persists, return for
service.
Wait until the unit discharges before
taking the next measurement.
Wait for more than 10 seconds after
power-up before taking blood pressure
measurements.
Measure patient's blood pressure with
other equipment.
Check cuff fit and positioning.
Switch cuff to other arm.
If problem persists, return for service.
Check cuff and hose.
Repeat NIBP measurement.
9650-0914-01 Rev. C