ZOLL R Series Manual page 26

Non-invasive blood pressure
Hide thumbs Also See for R Series:
Table of Contents

Advertisement

R S
— N
-I
ERIES
ON
NVASIVE
To stop automatic measurements, simply press the NIBP Auto softkey. (If the NIBP Auto
softkey label is not displayed, see "Displaying the NIBP Menu" on page 12.)
Aborting Measurements
You can immediately abort a blood pressure measurement at any time whether you are in
MONITOR, DEFIB or PACER mode (ON for BLS/Plus models, in Manual mode only).
To abort the current measurement:
Press the NIBP button (
The R Series unit immediately stops taking the measurement, displays the
NIBP MEAS ABORTED message, and deflates the cuff.
WARNING!
Make sure that a measurement has already started before attempting to abort. If you
press the NIBP button while a measurement is not being taken, the R Series unit starts a
new measurement.
NIBP Button
Note that:
If you press the NIBP button while taking a STAT measurement, the R Series unit
terminates all subsequent measurements.
If you press the NIBP button while the R Series unit is taking an automatic measurement, it
aborts the current measurement, deflates the cuff, and takes all subsequent measurements at
the set interval.
22
B
P
(NIBP) O
LOOD
RESSURE
) on the R Series front panel.
CPR
ECG
IDLE
SpO2%
0:00
x1
PADS
Depth
99
Rate
PPI
NIBP
mmHg
122
80
ECG
94
CO2
mmHg
---
---
RR
MONITOR
11:24
Report
Code
Options
Param
Alarms
Marker
Data
www.zoll.com
I
PTION
NSERT
LEAD
ANALYZE
ANALYZE
72
SIZE
ALARM
SUSPEND
RECORDER
MONITOR
OFF
PACER
4:1
OUTPUT
OUTPUT
mA
mA
3
SHOCK
SHOCK
2
CHARGE
CHARGE
ENERGY
ENERGY
SELECT
SELECT
1
DEFIB
RATE
RATE
ppm
ppm
9650-0914-01 Rev. C

Hide quick links:

Advertisement

Table of Contents
loading

Related Products for ZOLL R Series

Table of Contents