ZOLL R Series Manual page 7

Non-invasive blood pressure
Hide thumbs Also See for R Series:
Table of Contents

Advertisement

Do not use damaged cuffs, hoses, or connectors.
Ensure proper cuff selection and placement to avoid inaccurate measurements or patient
injury.
Position the cuff so it is level with the heart during measurement.
Do not attach the cuff to a limb being used for IV infusion, SpO
monitoring equipment. Cuff inflation might block the infusion, causing harm to the patient
or inaccurate SpO
Do not repeat NIBP measurements (particularly STAT measurements) at intervals less than
3-5 minutes over an extended period of time. Rapidly repeating measurement can impair
circulation in the monitored limb.
Do not sterilize or immerse the cuffs or hoses.
9650-0914-01 Rev. C
measurements.
2
R Series Non-Invasive Blood Pressure (NIBP)
Cuff and Hose Safety
monitoring, or other
2
3