Installing The Tube; Installing The Continuous Tube - VERDER Verderflex Vantage 5000 Series Original Operating Manual

Peristaltic cased tube pump
Table of Contents

Advertisement

5.2

Installing the Tube

DANGER
Isolate power supply from the pump before opening the
u
pump head.
CAUTION
Make sure the tube is compatible with the rotor assembly.
u
Before using a new tube assembly, make sure the pump is
u
run in the counter-clockwise direction for 1 minute.
5.2.1 Vantage 5000 Pump Tube Options
1. Continuous Tube
Figure 6
Vantage 5000 Pump Tube Options
Note
The tube element incorporates a drain port as show in Figure 6
(2. Tube Element) which must be utilised to prevent build-up of
pressurised fluid in the pump head in the event of tube failure.
Cut 1-2mm from the tip of the drain port and attach a sufficient
length of tubing into a bunded area or collection vessel.
Vantage 5000
2. Tube Element
1.4v-09.2019
14 | Page