Running the Pump Reagent task to empty the wash station
AssayMAP 96AM Wash Station Maintenance Guide
CAUTION
Using an incorrect profile can cause equipment damage. Ensure that the profile
is correct for the head type and deck configuration.
To empty the wash station using the Pump Reagent task:
1
At the carboy that contains the wash solution, remove the lid and lift it so
that the tubing is above the liquid level.
2
Ensure that the Protein Sample Prep Workbench is closed, and then start
the VWorks software.
You can use the VWorks desktop icon or the Start menu on the Windows
desktop to start the VWorks software.
3
Ensure that Simulation is off.
4
In the VWorks window, open the AssayMAP Bravo device file
(AssayMAP Bravo.dev).
To do this, click File > Open. In the Open dialog box, ensure the file type is
Device Files (*.dev), select the file, and then click Open.
The device files are installed in the following default location:
C:\VWorks Workspace\Workbench\Device Files
5
Click No when the Would you like to initialize devices now? message appears.
6
In the AssayMAP Bravo.dev tab, click Device diagnostics. The Bravo Diagnostics
dialog opens.
1 96AM Wash Station Maintenance Guide
Emptying the 96AM Wash Station