1 96AM Wash Station Maintenance Guide
Replacing damaged chimneys
12
4
Using your thumb, press down on the chimney so that the chimney touches
the bottom surface of the wash station and is fully seated.
You can insert a rubber pad between your thumb and the chimney to
protect your thumb.
5
Ensure that all the chimneys in the wash station are oriented so that the
dots are at the top end of the wash station. Ensure that the tops of the
chimneys are level with each other.
Figure Chimney dots oriented at top and tops of chimneys aligned
6
Repeat steps 1 to 5 to replace any other defective chimneys.