1 96AM Wash Station Maintenance Guide

Uninstalling the 96AM Wash Station

Uninstalling the 96AM Wash Station
10
WARNING
To prevent potential injury, turn off the Bravo Platform before you install or
remove any accessory.
To uninstall the 96AM Wash Station:
1
On the side of the Bravo Platform, press the power switch to the off (o)
position.
2
Gently, move the Bravo head manually to position it over deck location 6,
which will provide unobstructed access to the wash station at deck
location 1.
Figure AssayMAP Bravo Platform with head positioned at deck location 6
3
On the left side as you face the front of the AssayMAP Bravo Platform,
loosen the two 2- mm screws using a hex driver. Uninstall the bracket that
secures the wash station to the platepad.
Figure 96AM Wash Station secured with platepad brackets
4
On the opposite side of the platepad, guide the wash station horizontal slot
off the platepad inner bracket.
AssayMAP 96AM Wash Station Maintenance Guide