Agilent Technologies AssayMAP 96AM Maintenance Manual page 7

Wash station
Table of Contents

Advertisement

AssayMAP 96AM Wash Station Maintenance Guide
Perform the following steps in the order given.
Step
For this task...
1
Inspect the chimneys for damage and
irregularities.
2
Empty the wash station.
3
Uninstall the wash station.
4
Replace the damaged chimneys.
5
Install the wash station.
6
Verify the wash station teachpoint
and adjust it, as necessary:
a
Initialize the AssayMAP 1
profile and adjust the
teachpoint for the 96AM Wash
Station.
b
Use the AssayMAP Teachpoint
Update protocol to update the
teachpoint in the remaining
profiles in the Protein Sample
Prep Workbench
automatically.
7
Start up the AssayMAP Bravo
Platform and verify that all the
chimneys work consistently.
1 96AM Wash Station Maintenance Guide
Required resources and workflow
See...
"Determining if the
chimneys are in good
condition" on page 6
"Emptying the 96AM Wash
Station" on page 6
"Uninstalling the 96AM
Wash Station" on page 10
"Replacing damaged
chimneys" on page 11
"Installing the wash station
and connecting the tubing"
on page 13
AssayMAP Bravo Platform
Installation Guide
service personnel) or
Bravo Platform User Guide
"Testing the 96AM Wash
Station" on page 16