Replacing damaged chimneys

AssayMAP 96AM Wash Station Maintenance Guide
5
Lift the wash station off of the platepad.
For ease of handling, you may disconnect the tubing from the wash station
using the quick- disconnect fittings.
To replace the damaged chimneys:
1
Using needle- nose pliers, pull out the defective chimneys from the wash
station.
IMPORTANT
Ensure that you discard any chimneys that you remove.
Using pliers on a chimney will permanently damage the chimney.
2
Inspect the wash station hole in which the chimney was seated. Ensure
that the hole is empty and clean.
If necessary, clean the hole.
3
Orient the new plastic chimney so that the end with the little dot is at the
top, and then insert the chimney into the hole in the wash station.
Figure Dot on chimney oriented at top end for installation