QuickTake 15 Performance Profile
Flow range:
Flow compensation
accuracy:
Compensating flow
back pressure range:
Typical Back Pressure of Sampling Media (inches water)
Flow Rate (L/min)
Filter/Pore Size (µm)
37-mm MCE/0.8
37-mm PVC/5.0
Compare the information in this table to pump compensation range to determine appropriate applications.
Power:
Battery recharge time: Approximately 5 hours
Typical Run time:
Temperature:
Run time features:
Flow fault features:
5 to 15 L/min
± 5% of set flow
15 L/min at 8 inches water back pressure
10 L/min at 15 inches water back pressure
5 L/min at 20 inches water back pressure
5.0
8.0
11
18
4
7
Rechargeable 7.2-volt lithium-ion (Li-Ion) battery (see Li-Ion Battery
Shipment on page 4), or battery with standard AC charger/adapter
(115 V), or battery with deluxe AC charger/adapter (100-240 V)
VersaTrap Cassette: 4 hours at 15 L/min
Results when tested with a new pump and new fully charged
battery. Pump performance may vary.
The AC charger/adapter can be used to extend battery run
time, but it is not a battery eliminator. Therefore, it will
not provide indefinite run times.
Operating:
Storage:
Charging:
Protect sample pump from weather when in use outdoors.
User-selectable features, user-adjustable presets (see Advanced
Setup)
Preset timed runs:
Manually set continuous
run with manual shut-off:
Intermittent sampling:
If the pump is unable to compensate due to excessive back pressure,
the pump will go into flow fault mode (see page 12 for details). Flow
fault features are enabled as the default setting. Flow fault may be
disabled at the user's option. See Advanced Setup.
Flow fault:
Fault restart:
10.0
12.0
22
28
9
11
32 - 113 F (0 - 45 C)
-4 - 113 F (-20 - 45 C)
32 - 113 F (0 - 45 C)
1, 2, 5, 10, or 15 min
1 to 999 min (repeats 1 to 999-min
runs indefinitely until user stops
pump or power supply is depleted)
See Advanced Setup
Flow fault displays immediately.
Pump shuts off after 5 seconds
Pump attempts restart every
10 seconds up to five times

15.0
36
15