• To navigate presets and displays:
 or 
Press
• To select a run time preset and run the pump:
Scroll to desired run time preset.
Select preset by pressing []. The pump will start to run.
Selecting the
Available Presets
Press  or 
• To run the pump:
Press [].
• To place the pump in hold:
Press [].
• To repeat a sample run:
From Done, reset (rSEt), or sampling error (SErr), press [] to return to
run time preset. Press [] to sample.
• To turn off LED:
Pump automatically turns off LED after it is idle for 4 minutes. User can
reactivate the LED by pressing and holding
• To interrupt run, terminate a sample, or reset pump:
From Hold, press
pump or
to ignore reset and return to Hold.
• To set flow or calibrate:
See Calibration and Setting Flow for details.
1. Connect a flowmeter to the pump inlet.
2. Scroll to any run time preset. Press and hold CAL for 2 seconds to enter
calibration mode (pump will start running).
3. Press  or  to change flow rate until desired flow appears on the
calibrator (Flow rate does not display on LED).
4. Press  and Stor will display. Press [] to save setting or
changes.
to scroll through presets or displays.
Press 
. Sample reset (rSEt) will display. Press [] to reset
.

Press 
15
10
5
2
1
Intt
Cont
to ignore