7.2. CALIBRATION MODE

■ Due to characteristics of the gas detector, minimum 30min of stabilization time is required and maintenance
condition may change depending on site condition.

7.2.1. Zero Calibration

- Contacting "FUNC" key with the Magnet-bar for 2 sec or longer in Measuring Mode enters Password
PASSWORD
mode.
[**]
- After setting Password using "↑" or "↓" key, contact "FUNC" key.
- Contact "↑" or "↓" key to select "Calibration Mode".
CALIBRA.
- Contact "FUNC" key when "CALIBRA. MODE" is displayed to enter Calibration Mode.
MODE
- Contact "RESET" key to return to Measuring Mode.
CALIBRA.
- When "CALIBRATION MODE" is displayed, contacting "FUNC" key selects Zero Calibration.
[ZERO]
- Contact "↑" or "↓" key to achieve [ZERO] then contact "FUNC" key to enter Zero Calibration.
ZERO CAL
- Contact "↑" or "↓" key to achieve [YES] then contact "FUNC" key to perform Zero Calibration.
[ NO]
ZERO GAS
- Using a calibration tool, inject clean air or 100% nitrogen into the sensor at a flow rate of 500 mL/min
[
0]
for 1 min. Contact "FUNC" key when measurement is stabilized to automatically perform Zero Calibration.
ZERO >>
- When zero calibration is successful, "ZERO SUCCESS" is displayed for 2 sec on LCD display then it
[SUCESS]
changes to "CALIBRATION DATA" Mode.
- When zero calibration is not successful, "ZERO FAIL" is displayed for 2 sec and it changes to
ZERO >>
"Calibration Data Mode".
[ FAIL ]
- ZERO FAIL occurs when sensor input value exceeds 70% of total ADC input range.
- It is a mode to display measurement after calibration and checks whether calibration is successful after
CAL.DATA
performing auto calibration.
[
0]
- When it failed, it displays "FAIL" and the current measurement in 1 sec interval.
- Contact"RESET" key to return to "Calibration mode".