4
Monitoring
4.6
SpO
-
SpCO
SpMet monitoring (Option)
2
,
,
Alarm (measurement field)
Check sensor
Page 54
Code
Cause
T.SP216
• No sensor connected
T.SP217
• Sensor life expired
T.SP218
• Incompatible sensor
T.SP219
• Unrecognised sensor
T.SP220
• Cable and sensor fault
T.SP221
• No adhesive sensor connected
T.SP222
• Adhesive sensor life expired
T.SP223
• Incompatible adhesive sensor
T.SP224
• Unrecognised adhesive sensor
T.SP225
• Defective adhesive sensor
T.SP227
• Defective sensor
DEFIGARD/PHYSIOGARD Touch 7
Remedy
 Check/replace the sensor