12
Technical Data
12.1 System data
Environmental conditions
For transient operation
Environmental conditions
For Transport and storage
between uses
Time for warming up/cooling
down
MIL STD 810G
Environmental conditions
Defibrillation electrodes
Storage
Storage max. 10 days
Display
Type
Dimensions
Alarm sound level
Connections
Interfaces
Page 132
Operation in NORMAL USE for a period not more than 20 min under the
following environmental operating conditions:
• a temperature range of - 20 °C to + 50 °C;
• a relative humidity range of 15 % to 90 %, non-condensing, but not requiring a
water vapour partial pressure greater than 50 hPa.
Operation in NORMAL USE for a period of 1 hour under a temperature range of
- 10°C to + 50°C;
• - 40 °C to + 5 °C without relative humidity control;
• + 5 °C to + 35 °C at a relative humidity up to 90 %, non-condensing;
• > 35 °C to 70 °C at a water vapour pressure up to 50 hPa;
after having been removed from its protective packaging and subsequently
between uses.
• 30 minutes;
Time required for the DEFIGARD/PHYSIOGARD Touch7 to warm or cool from
the minimum/maximum storage temperature between uses until the DEFIGARD/
PHYSIOGARD Touch7 is ready for its intended use when the ambient
temperature is 20 °C.
Device was tested in accordance with:
• MIL STD 810G Method 501.5 Procedure III
• Tactical-standby to Operational (+60°C)
• MIL STD 810G Method 502.5 Procedure II Operation (-26°C).
• 0 ºC...50 ºC
• -40 ºC...75 ºC
• High-resolution colour LCD capacitive touch screen, protected by tempered glass
•
7 " (154 x 85.92 mm)
65 dBA for medium and high priority alarms
ECG patient cable, SpO
, NIBP, Temperature, CO2, IBP
2
USB
DEFIGARD/PHYSIOGARD Touch 7