DEFIGARD/PHYSIOGARD Touch 7
Step 3
Step 4
User guide
Step 3 shock delivery
As soon as the energy for a shock is charged, the device prompts the user to deliver
the shock by pressing button 3.
 Danger of electric shock!
• Do not, under any circumstances, touch the patient during shock delivery.
• Make sure that the patient does not touch any conducting objects.
9. Deliver the shock by pressing the button
After the shock delivery, step 4 follows.
The following default energy values are programmed:
1
2
3
Cardiopulmonary resuscitation
10. Carry out cardiopulmonary resuscitation. Alternate between 30 chest compres-
1
sions and 2 breaths
for 2 minutes
Step 2,
Analysis.
11. Finish the therapy (see page
The CPR duration may vary according to country-specific standards (see page
Defibrillator ERC Protocol).
1. A "continuous compressions" option is also available (i.e. no rescue breaths)
2. CPR cycle duration can vary depending on the "CPR cycle configuration" settings.
Semi-automated defibrillation
Shock
Adults
150 joules
200 joules
200 joules
2
. After 2 minutes, the device begins again with
106).
Defibrillation
5
5.5
.
Children
50 joules
50 joules
50 joules
110
Page 95