DEFIGARD/PHYSIOGARD Touch 7
Standard energy settings
AED
Adult
Paediatric
Manual mode
Adult
Paediatric
Charging time for shock
• with fully charged battery
• with 15 VDC mains voltage after 15
discharges with max. energy
• from switch-on of the device with
pads
Cycle Time Rhythm Analysis –
Shock Standby in AED Mode
• with fully charged battery
• with 15 VDC mains voltage after 15
discharges with max. energy
• from switch-on of the device to
charge at max. energy
Cycle time shock – shock
Operating Modes
Charge control and monitoring
Patient resistance
Indication when ready to shock
Shock delivery
User guide
Deviation at 50 3 J or 15 % (the higher value is assumed)
150/200/200 joules (configurable see page 149)
50/50/50 joules (configurable see page 149)
(automatic selected when the paediatric or adult electrodes are connected)
150 joules (configurable see page 149)
50 joules (configurable see page 149)
(default energy settings when starting in manual mode, adjustable anytime during
intervention)
(Time used to charge the storage capacitor to the max. energy of 200 J in manual
mode)
8 seconds
9 seconds
19 seconds
1st shock = 11 s max.
1st shock = 11 s max.
1st shock = 23 s
<15 s
• Synchronised with heart action < 60 ms after R wave
• Non synchronised
• AED
• Automatic shock recommendation of analysis in AED mode
• Direct via touch screen
• Display of selected energy
25 ...250 
LED below
is lit
Using key
Technical Data
Defibrillation Waveform
Page 135