Routing A Signal To Counter N Gate; Routing Counter N Gate To An Output Terminal; Counter N Aux Signal; Routing A Signal To Counter N Aux - National Instruments M Series User Manual

Multifunction i/o modules and devices
Hide thumbs Also See for M Series:
Table of Contents

Advertisement

Chapter 7
Counters

Routing a Signal to Counter n Gate

Each counter has independent input selectors for the Counter n Gate signal. Any of the following
signals can be routed to the Counter n Gate input:
RTSI <0..7>
PFI <0..15>
AI Reference Trigger (ai/ReferenceTrigger)
AI Start Trigger (ai/StartTrigger)
AI Sample Clock (ai/SampleClock)
AI Convert Clock (ai/ConvertClock)
AO Sample Clock (ao/SampleClock)
DI Sample Clock (di/SampleClock)
DO Sample Clock (do/SampleClock)
PXI_STAR
Change Detection Event
Analog Comparison Event
In addition, Counter 1 Internal Output or Counter 1 Source can be routed to Counter 0 Gate.
Counter 0 Internal Output or Counter 0 Source can be routed to Counter 1 Gate.
Some of these options may not be available in some driver software.

Routing Counter n Gate to an Output Terminal

You can route Counter n Gate out to any PFI <0..15> or RTSI <0..7> terminal. All PFIs are set
to high-impedance at startup.

Counter n Aux Signal

The Counter n Aux signal indicates the first edge in a two-signal edge-separation measurement.

Routing a Signal to Counter n Aux

Each counter has independent input selectors for the Counter n Aux signal. Any of the following
signals can be routed to the Counter n Aux input:
RTSI <0..7>
PFI <0..15>
AI Reference Trigger (ai/ReferenceTrigger)
AI Start Trigger (ai/StartTrigger)
PXI_STAR
Analog Comparison Event
7-26 | ni.com