Program 5 has a double cycle making it possible to set two independent temperatures at the same time.
Program 6 can be used to run a cycle with the additional information of the F0 parameter. The
calculation of this parameter is based on the setting selected in sub menu F0 in the option menu.
Program 7 can be used to run a cycle at the selected temperature until the selected value of the F0 parameter
is reached. The calculation of this parameter is based on the setting selected in sub menu F0 in the option
menu.

Configuration parameter settings

Press the MENU button to modify the instrument configuration parameter settings.
There are eight sub menus:
• "Printer" - modifies the printer settings
• "F0" - sets the value for F0 calculation on programmes 6 and 7
• "Delay" - sets a delayed start of the sterilisation cycle
• "Print last cycle" - prints the last sterilisation cycle completed
• "Password" - sets a password to prevent unauthorised modification of the sterilisation and instrument
settings
• "Date and time" - modifies the autoclave date and time
• "LCD contrast" - modifies the adjustment of the display contrast
• "Service" - This menu is for maintenance authorised personnel only. The parameters stored in this menu can
only be displayed but not modified...with the exception of the language.
Sub menu "Printer"
▲ P R
From this menu it is possible to enable or disable the printer plus set two different printing intervals for the
sterilisation and heating/cooling phases. Press the ENTER button from the "PRINTER" menu; the workflow is
as follows:
I
N T E R
11