Detector Temperature, Acquire, And Focus Are Grayed Out (Versions 2.5.19.0 And Earlier); Detector Temperature, Acquire, And Focus Are Grayed Out; (Versions 2.5.19.0 And Earlier) - Princeton Instruments PhotonMAX System Manual

Table of Contents

Advertisement

74
5. Because the PhotonMAX camera is second in the list (PM Style), you will now

Detector Temperature, Acquire, and Focus are Grayed Out

(Versions 2.5.19.0 and earlier)

These functions and others will be deactivated if you have installed a camera being run
under PVCAM and have opened WinView/32 or WinSpec/32 without having first turned
on the PhotonMAX camera. They will also be deactivated if you have installed a camera
being run PVCAM and a Princeton Instruments (RSPI) high speed PCI card was also
detected when RSConfig.exe was run.
1. Check to see if the PhotonMAX is connected to the host computer and is
2. Close WinView, verify that the PhotonMAX is connected to the host computer,
3. If the PhotonMAX camera (PM Style) is connected and powered on, it may not
4. Run RSConfig.exe (accessible from the Windows|Start|Programs|Roper
PhotonMAX System Manual
Figure 62. RSConfig dialog box: Two Camera Styles
have to edit the PVCAM.INI file created by RSConfig.exe. Refer to "Editing
the PVCAM.INI File", page 75, for instructions.
powered on. If it is not connected or is connected but not powered on, go to Step
2. If it is connected and on, go to Step 3.
power the PhotonMAX on, and reopen WinView. The formerly grayed out
functions should now be available.
be listed as Camera 1 in the PVCAM.INI file.
Scientific menu). If the PhotonMAX camera is listed as Camera 2 (PM Style) in
Figure 63, you will have to edit the PVCAM.INI file. Refer to "Editing the
PVCAM.INI File", page 75, for instructions.