56
Bulb Trigger
In Bulb Trigger mode, exposure time for each frame is determined by the trigger pulse
width. Exposure time entered into the software is ignored in this mode (see Figure 43). If
a trigger arrives during the readout of the previous frame, it is ignored. The shaded areas
denote the idle time between exposures.
Trigger In
Exposure
Frame Readout
Single Trigger
In this mode, the camera requires only one trigger to acquire a sequence of frames. Once
the initial trigger is received, the camera ignores any further triggers until the entire
exposure/readout sequence is completed (see Figure 44 and Figure 45, which show a
three-frame sequence).
PhotonMAX System Manual
Ignored
#1
#1

Figure 43. Bulb Mode Timing Diagram: Non-Overlap Mode

Trigger In
Clears
#1
Exposure
Frame Readout
Figure 44. Single Trigger Mode Timing Diagram: Overlap Mode
Trigger In
Clears
#1
Exposure
#1
Frame Readout
Figure 45. Single Trigger Mode Timing Diagram: Non-Overlap Mode
#2
#2
Ignored
#2
#3
#1
#2
Ignored
#2
#2
Version 1.B
#3
#3
#3
#3