Figure 16. Cleans Tab - Princeton Instruments PhotonMAX System Manual

Table of Contents

Advertisement

32
PhotonMAX System Manual
Detector Temperature (Setup|Detector Temperature...): The default
temperature setting is read from the camera. When the array temperature
reaches the set temperature, the Detector Temperature dialog box will
report that the temperature is LOCKED . Note that some overshoot may occur.
This could cause temperature lock to be briefly lost and then quickly re-
established. If you are reading the actual temperature reported by the
application software, there may be a small difference between the set and
reported temperature when lock is established. This is normal and does not
indicate a system malfunction. Once lock is established, the temperature will
be stable to within ±0.05°C.
Figure 17. Detector Temperature dialog box
Note: The Detector Temperature dialog box will not display temperature
information while you are acquiring data.
Experiment Setup|Main tab page (Acquisition|Experiment Setup...):
Exposure Time: 100 ms
Accumulations & Number of Images: 1
Experiment Setup|ROI tab page (Acquisition|Experiment Setup...):
Use this function to define the region of interest (ROI).
Imaging Mode: Select this mode if you are running WinSpec.
Select
"Clear Pre-Sequence"
Figure 16. Cleans tab page