Figure 15. Controller/Camera Tab - Princeton Instruments PhotonMAX System Manual

Table of Contents

Advertisement

Chapter 5
Setting the Parameters
Note: The following procedure is based on WinView/32: you will need to modify it
if you are using a different application. Basic familiarity with the WinView/32
software is assumed. If this is not the case, you may want to review the software
manual or have it available while performing this procedure.
Set the software parameters as follows:
Controller/Camera tab page (Setup|Hardware): These parameters should
be set automatically to the proper values for your system.
Use PVCAM: Verify that this box is checked.
Note: This check box is not present on software versions 2.5.19.6 and
higher.
Controller type: This information is read from the camera.
Camera type: This information is read from the camera.
Shutter type: Custom (System dependent).
Readout mode: Available modes are read from the camera. Select
Frame Transfer.
Figure 15. Controller/Camera tab page
Cleans tab page (Setup|Hardware): Select Clear Pre-Sequence as the
Clean Mode.
Operation
Select the
"Frame Transfer"
Readout Mode
31