PhotonMAX System Manual
Controller type: This information is read from the camera.
Camera type: This information is read from the camera.
Shutter type: Custom (System dependent).
Readout mode: Available modes are read from the camera. Select
Frame Transfer.
Figure 19. Controller/Camera tab page
Detector Temperature (Setup|Detector Temperature...): The default
temperature setting is read from the camera. When the array temperature
reaches the set temperature, the Detector Temperature dialog box will
report that the temperature is LOCKED . Note that some overshoot may
occur. This could cause temperature lock to be briefly lost and then quickly
re-established. If you are reading the actual temperature reported by the
application software, there may be a small difference between the set and
reported temperature when lock is established. This is normal and does not
indicate a system malfunction. Once lock is established, the temperature will
be stable to within ±0.05°C.
Figure 20. Detector Temperature dialog box
Note: The Detector Temperature dialog box will not display temperature
information while you are acquiring data.
Select the
"Frame Transfer"
Readout Mode
Version 1.B