CONTINUED...
Recommended Separation Distances between Portable and Mobile RF Communications Equipment
and the Prima ENT/DNT for ME Equipment and ME systems that are NOT Life-supporting.
Guidance and Manufacturer's Declaration -Electromagnetic Immunity
Portable and Mobile RF Communications Equipment and the Prima ENT/DNT
The Prima ENT/DNT is intended for use in electromagnetic environments in which radiated
RF disturbances are controlled. The customer or user of the Prima ENT/DNT can help prevent
electromagnetic interference by maintaining a minimum distance between portable and mobile
RF Communications Equipment (transmitters) and the Prima ENT/DNT as recommended below,
according to the maximum output power of the communications equipment.
Max Output Power
of Transmitter(W)
0.01
0.1
1
10
100
For transmitters rated at a maximum output power not listed above, the recommended separation
distance (d) in meters(m) can be estimated using the equation applicable to the frequency of the
transmitter, where P is the maximum output power rating of the transmitter in watts(w) according to
Note 1: At 80 MHZ, the separation distance for the higher frequency range applies.
Note 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects, and people.
36
Recommended Separation Distances for between
Separation
Separation(m)
(m) 150kHz
80 to 800 MHZ
to 80 MHZ
d=(3.5/E1)(Sqrt P)
d=(3.5/v1)
(Sqrt P)
0.1166
0.1166
0.3689
0.3689
1.1666
1.1666
3.6893
3.6893
11.6666
11.6666
the transmitter manufacturer.
Prima ENT/DNT
Separation (m) 800 MHZ to 2.5Ghz d=(7/E1)
(Sqrt P)
0.2333
0.7378
2.3333
7.3786
23.3333