Agilent Technologies U1602A User's And Service Manual page 74

Handheld digital oscilloscopes
Hide thumbs Also See for U1602A:
Table of Contents

Advertisement

3
Using the Scope Functions
74
• Slope
Press F3 on page 1/2 to select rising edge
The selected slope is displayed at the trigger status line at the right
of the display.
• Coupling
Press F2 on page 2/2 to select trigger coupling:
DC coupling allows DC and AC signals to pass through the trigger
path.
AC coupling allows AC signals to pass through the trigger path
and filter DC offset voltage from the trigger waveform. The
coupling cutoff frequency is less than 1 Hz. The AC coupling is
useful to get a stable edge trigger when the waveform has a large
DC offset.
HF Rejection coupling places a 50 kHz low- pass filter in the
trigger path to remove high frequency components from the trigger
waveform. The HF- Rej is useful to observe low frequency signals
while getting rid of high frequency noise from trigger path.
LF Rejection coupling places a 30 kHz high- pass filter in series
with the trigger waveform to remove any unwanted low frequency
components from the trigger waveform. This LF- Rej is useful to
obtain a stable edge trigger when your waveform contains low
frequency noise.
Noise Rejection coupling is three times the DC coupled limits. The
Noise- Rej provides additional hysteresis to the trigger circuitry.
• Level
Press F3 on page 2/2 to select the types of trigger level adjustment.
Manual trigger level allows you to change the trigger level by
turning the rotary switch. The value of the trigger level is
displayed under the Level column.
TTL trigger level sets the trigger level to recognize the High/Low
condition in a TTL signal automatically.
Agilent U1602A/U1604A User's and Service Guide
or falling edge
.