Resistance Verification Test (2-Wire Ohms)
Calibrator Settings
600
W
6 k
W
60 k
W
600 k
W
3 M
W
6 M
W
60 M
W
Capacitance Verification Test
Calibrator Settings
60 nF
600 nF
6
F
m
36
F
m
60
F
m
300
F
m
Diode Verification Test
Calibrator Voltage
0.9 V
Continuity Verification Test
Calibrator Settings
30
W
75
W
Agilent U1602A/U1604A User's and Service Guide
Range
Test Limit
600
595.8
W
6 k
5.958 k
W
60 k
59.58 k
W
600 k
595.8 k
W
6 M
2.973 M
W
6 M
5.958 M
W
60 M
59.28 M
W
Range
Test Limit
60 nF
58.68 nF to 61.32 nF
600 nF
586.8 nF to 613.2 nF
6
F
5.868
m
60
F
35.16
m
60
F
58.68
m
300
F
293.4
m
Test Limit
0.8812 V to 0.9188 V
Condition
Buzzer sounds on
Buzzer sounds off
to 604.2
W
W
to 6.042 k
W
W
to 60.42 k
W
W
to 604.2 k
W
W
to 3.027 M
W
W
to 6.042 M
W
W
to 60.72 M
W
W
F to 6.132
F
m
m
F to 36.84
F
m
m
F to 61.32
F
m
m
F to 306.6
F
m
m
Results
Results
[PASS/FAIL]
[PASS/FAIL]
Performance Tests
Results
Results