Agilent Technologies U1602A User's And Service Manual page 198

Handheld digital oscilloscopes
Hide thumbs Also See for U1602A:
Table of Contents

Advertisement

9
Characteristics and Specifications
DIGITAL MULTIMETER SPECIFICATIONS
Function
DC Voltage
AC Voltage
AC+DC Voltage
Resistance
Capacitance
Diode
[1] For temperature between 0
198
[1]
Range
600.0 mV
6.000 V
60.00 V
600.0 V
600.0 mV – 600.0 V
6.0000 V – 600.0 V
600.0
W
6.000 k
W
60.00 k
W
600.0 k
W
6.000 M
W
60.00 M
W
60.00 nF
600.0 nF
6000 nF
60.00 μF
300.0 μF
1.000 V
for every degree Celsius.
± (% of reading + % of range)
Frequency, Test Current or
Burden Voltage
50 Hz – 1 kHz
1 kHz – 30 kHz
50 Hz – 1 kHz
1 kHz – 30 kHz
0.5 mA
o
o
o
C to 18
C and 28
C to 50
Agilent U1602A/U1604A User's and Service Guide
1 year Tcal ± 5
0.3 + 0.08
0.3 + 0.08
0.3 + 0.08
0.3 + 0.08
1.0 + 0.2
3.0 + 0.2
1.0 + 0.2
3.0 + 0.2
0.5 + 0.2
0.5 + 0.2
0.5 + 0.2
0.5 + 0.2
0.5 + 0.2
1.0 + 0.2
2.0 + 0.2
2.0 + 0.2
2.0 + 0.2
2.0 + 0.2
2.0 + 0.2
2.0 + 0.08
o
C , add 0.1% of reading + 0.02% of range
o
C

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

U1604a