Agilent Technologies U1602A User's And Service Manual page 180

Handheld digital oscilloscopes
Hide thumbs Also See for U1602A:
Table of Contents

Advertisement

7
Performance Tests
U1604A
Frequency: 10 kHz
Period: 100
Frequency: 1 MHz
Period: 1
Frequency: 40 MHz
Period: 25 ns
Trigger Sensitivity
Calibrator Setting
100 mVp-p at 5 MHz
100 mVp-p at 20 MHz
100 mVp-p at 40 MHz
Multimeter Performance Verification Tests
DC Voltage Verification Test
Calibrator Voltage
600 mV
6 V
30 V
60 V
600 V
AC Voltage Verification Test
Calibrator Voltage
6 V at 1 kHz
6 V at 30 kHz
60 V at 1 kHz
60 V at 30 kHz
300 V at 50 Hz
540 V at 1 kHz
590 V at 10 kHz
180
9.63 kHz to 10.4 kHz
s
96.20
m
962 kHz to 1.04 MHz
s
0.96
m
38.24 MHz to 41.93 MHz
23.85 ns to 26.15 ns
Test Limits
0.8 division at 5 MHz
1 division at 20 MHz (U1602A)
1 division at 40 MHz (U1604A)
Range
600 mV
6 V
60 V
60 V
600 V
Range
6 V
6 V
60 V
60 V
600 V
600 V
600 V
s to 103.8
s
m
m
s to 1.04
s
m
m
Test Limit
597.7 mV to 602.3 mV
5.977 V to 6.023 V
29.86 V to 30.14 V
59.77 V to 60.23 V
597.7 V to 602.3 V
Test Limit
5.928 V to 6.072 V
5.808 V to 6.192 V
59.28 V to 60.72 V
58.08 V to 61.92 V
295.80 V to 304.20 V
533.40 V to 546.60 V
571.1 V to 608.9 V
Agilent U1602A/U1604A User's and Service Guide
Channel 1
Channel 2
[PASS/FAIL]
[PASS/FAIL]
[PASS/FAIL]
[PASS/FAIL]
[PASS/FAIL]
[PASS/FAIL]
Results
Results