Agilent Technologies U1602A User's And Service Manual page 167

Handheld digital oscilloscopes
Hide thumbs Also See for U1602A:
Table of Contents

Advertisement

1 Connect the calibrator to channel 1 using a 50
2 Verify the trigger sensitivity at 5 MHz, 0.8 division.
3 Verify the trigger sensitivity at maximum bandwidth at 1 division.
4 Repeat the same verification procedures for channels 2.
Agilent U1602A/U1604A User's and Service Guide
oscilloscope input.
• Set the calibrator to "level sine" and OPR/STBY to "OPR" mode.
• Set output of the calibrator to 5 MHz and the amplitude to 100
mVp- p.
• Press the Autoscale key.
• Set the time base to 50 ns/div.
• Set channel 1 to 100 mV/div.
• Decrease the output signal of the calibrator until 0.8 vertical division
of the signal is displayed.
• The trigger is stable when the displayed waveform is stable. If the
trigger is not stable, try to adjust the trigger level. If adjustment is
able to stabilize the waveform, the test result is still considered a
pass.
• Change the output signal of the calibrator to 40 MHz for the U1604A
or 20 MHz for the U1602A, and set the amplitude to 100 mVp- p.
• Press the autoscale key.
• Set the time base to 10 ns/div.
• Decrease the output of the calibrator until 1 vertical division of the
signal is displayed.
• The trigger is stable when the displayed waveform is stable. If the
trigger is not stable, try to adjust the trigger level. If the adjusment
able to stabilize the waveform, the test count is still considered a
pass.
Performance Tests
feedthrough at the
W
7
167