Philips M1013A IntelliVue G1 Service Manual page 33

Gas analyzer
Table of Contents

Advertisement

Uploading the Software
3 Software Uploads
2) Next a pop-up window will open. Click on the Select button and choose the first software file from
the folder to which it was downloaded. Both software files must be loaded to the device (except for
M1013A without O
option). The order in which the files are loaded is irrelevant. The Baud rate is
2
automatically set to 57600. Do not change this setting.
3) Click on the Start button and wait for the automatic software load procedure to begin. This may
take a few seconds (up to 15 sec to start) and can be observed in the status bar in the lower half of the
pop-up window. Do not interrupt this procedure.
The whole procedure can take up to 5 minutes for each file, but will normally require less than that.
33

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

M1019a intellivue g5

Table of Contents