Disposal Of Empty Gas Cylinder; Mounting Integrity Test; Reporting Of Test Results - Philips M1013A IntelliVue G1 Service Manual

Gas analyzer
Table of Contents

Advertisement

Reporting of Test Results

Disposal of Empty Gas Cylinder

Empty cylinder completely by pushing in the pin of the valve.
1
Once the cylinder is empty, drill a hole in the cylinder
2
Be careful to assure that the cylinder is completely empty before you try to drill the cylinder.
CAUTION
Write "Empty" on the cylinder and place it with your scrap metal or, if you do not collect scrap
3
metal for recycling, dispose of the cylinder.

Mounting Integrity Test

If at any time:
• one or more of the four of the quick mount screws are loose
• there is a clearance between the quick mount and the monitor bottom housing
• the G1/G5 mounting is unstable
remove the G1/G5 from the mount and disassemble the quick mount. Ensure that the threading of the
G1/G5 is not damaged.
If the quick mount is damaged, exchange the quick mount.
Ensure that all quick mount screws are tight (3.5 Nm). Test the quick mount by pressing the quick
release button. If it comes back out gradually and regularly, the quick mount is inserted correctly. If it
gets stuck, the quick mount is not centered and must be reinserted correctly.
If you notice any damage to the threading of the G1/G5 chassis, send the G1/G5 in for bench repair.
Test
Mounting Integrity Test
Reporting of Test Results
Philips recommends all test results are documented in accordance with local laws. Authorized Philips
personnel report the test result back to Philips. While hospital personnel (biomedical engineers or
technicians) do not need to report results to Philips, Philips recommends that they record and store the
test results in accordance with local laws.
The following table lists what to record after completing the tests in this chapter. Record the results in
the empty column in the Test and Inspection Matrix.
The following is a guide as to what your documentation should include:
• Identification of the testing body (for example, which company or department carried out the tests).
• Name of the person(s) who performed the tests and the concluding evaluation.
• Identification of the device(s) and accessories being tested (serial number, etc.).
Expected test results
All quick mount screws are tight. No
damage to quick or fix mount. No
damage to threading of G1/G5.
4 Testing and Maintenance
69