Repairing The Gas Analyzer; Introduction; Who Should Perform Repairs; Tools Required - Philips M1013A IntelliVue G1 Service Manual

Gas analyzer
Table of Contents

Advertisement

Repairing the Gas Analyzer

6

Introduction

This section contains detailed removal and replacement procedures for all field-replaceable units in the
gas analyzer.

Who Should Perform Repairs

Only qualified service personnel should open the IntellIVue G1 / G5 housing, remove and replace
components, or make adjustments. If your medical facility does not have qualified service personnel,
contact Philips' Response Center or your local Philips representative.
• Switch off the instrument and disconnect it from the mains power supply. Take standard
WARNING
electrostatic precautions. For example, a wrist strap connected to electrical ground.
• Voltages dangerous to life are present in the instrument. Do not perform any disassembly or
reassembly procedures with power applied to the instrument. Failure to adhere to this warning could
cause serious injury or death.

Tools required

• 1 hexagon socket screw key (4.0)
• 1 hexagon socket screw key (2.5)
• ESD mat and wrist strap
Your IntelliVue G1 / G5 may look slightly different than on the pictures in this chapter, depending on
N O T E
the options ordered
6