Safety Test Procedures - Philips M1013A IntelliVue G1 Service Manual

Gas analyzer
Table of Contents

Advertisement

Safety Tests
• For Europe and Asia/Pacific, the monitor complies with:
IEC 60601-1:1988 + A1:1991 + A2:1995; EN60601-1:1990 + A1:1993 + A2:1995; IEC 60601-1-
1:2001; EN 60601-1-1:2001; IEC 60601-1-2:2001; EN 60601-1-2:2001.
For USA, the monitor complies with:
UL60601-1
For Canada, CAN/CSA C22.2#601.1-M90
• Local regulations supersede the testing requirements listed in this chapter.
• If a non-medical electrical device is connected to a medical electrical device, the resulting medical
electrical system must comply with IEC/EN 60601-1-1.
• Perform safety tests as described on the following pages.

Safety Test Procedures

Use the test procedures outlined here only for verifying and recording the initial values prior to or at
installation, safe installation or service of the product, and for periodic recurrent testing. The setups
used for these tests and the acceptable ranges of values are derived from local and international
standards but may not be equivalent. These tests are not a substitute for local safety testing where it is
required for an installation or a service event. If using an approved safety tester, perform the tests in
accordance with the information provided by the manufacturer of the tester and in accordance with
your local regulations, for example IEC/EN 60601-1, UL60601-1 (US), IEC/EN 62353, and IEC/EN
60601-1-1. The safety tester should print results as detailed in this chapter, together with other data.
Please refer to Annex C of IEC/EN 62353 for requirements for the measurement equipment and for
measurement circuits for protective earth resistance and leakage currents.
4 Testing and Maintenance
45