Setup and Configuration Procedures

Setup and Configuration Procedures
This section describes final setting up and configuration procedures that must be completed after the
gas analyzer is connected to the monitor and switched on before the gas analyzer is used for
monitoring.

IntelliVue Serial Port Configuration

The MIB port used in the IntelliVue host monitor must be configured to "GM".
To do this, go into service mode and then select Setup followed by Hardware and then MIB.

Altitude Configuration

The altitude setting for the monitor is important as it is used as a reference to check the gas analyzer
ambient pressure measurement.
See your monitor service guide for details.

Connect Sample Input Tubing

Connect the sample input tubing to the watertrap at the luer lock connector. For details, refer to the
Instructions for Use.

Post-Installation Checks

See Test and Inspection Matrix for details.
Do not use the instrument for any monitoring procedure on a patient if you identify anything which
WARNING
indicates impaired functioning of the instrument.
Safety Requirements Compliance and
Considerations
The M1013A IntelliVue G1 and the M1019A IntelliVue G5 comply with the following international
safety requirements for medical electrical equipment:
IEC 60601-1:1988 + A1:1991 + A2:1995; EN60601-1:1990 + A1:1993 + A2:1995; UL 60601-
1:2003; CAN/CSA C22.2#601.1-M90; IEC 60601-1-2:2001; EN 60601-1-2:2001.
Classification (according to IEC 60601-1): Class 1, Type BF, Continuous Operation.
This ISM device complies with Canadian ICES-001. Cet appareil est conforme a la norme NMB-001
du Canada.
The possibility of hazards arising from software errors was minimized in compliance with
ISO14971:2000, EN60601-1-4:1996 + A1:1999 and IEC 60601-1-4:1996 + A1:1999.
2 Installation and Patient Safety