Software Uploads - Philips M1013A IntelliVue G1 Service Manual

Gas analyzer
Table of Contents

Advertisement

3

Software Uploads

3
For a software update of the M1013A IntelliVue G1 and M1019A IntelliVue G5 you need the most
recent software as listed in the Service Bulletin (SB) or Field Change Order (FCO) and download it
from the SoftServer as described there. Save it to a folder where it is easy to find, for example
C:/TEMP. There will be two software files, e.g. pg01400.img for software revision 1.40 and Multi
Function Module software such as mf030900.img. For M1013A with O
option and M1019A, both
2
files need to be downloaded and later loaded to the device. For M1013A without O
option only the
2
general software file beginning with "pg" is required; there is no MFM software for M1013A without
O
option, so there is no need for loading the file.
2
Please use the support tool VISIA 1.14.FF or higher and connect the M1013A / M1019A. The
connection is made automatically when VISIA is opened and the cables are plugged into the PC/
Laptop and the M1013A / M1019A; otherwise please press the "Connect" button in the GENERAL
Tab in the VISIA tool.
For automatic connection please power on the M1013A / M1019A 15 to 20 seconds before initializing
the VISIA tool.
27