Environmental Specifications; Mdd Classification; Performance Specifications - Philips M1013A IntelliVue G1 Service Manual

Gas analyzer
Table of Contents

Advertisement

Environmental Specifications

Environmental Specifications
Operating Temperature:
Storage Temperature:
Humidity Limit (Operating):
Humidity Limit (Storage):
Altitude Range (Operating):
Altitude Range (Storage):
Warm-up Time:

MDD Classification

According to the Council Directive 93/42/EEC (Medical Devices Directive) the device classification is
2A, Rule 10.

Performance Specifications

All Performance and accuracy specifications are valid based on gas sample tubing M1658A, including
watertrap M1657B, and airway adapter 13902A.
Humidity Correction: For CO
Wet: p [mmHg] = c [Vol%] * (p_abs - p_H
Dry: p [mmHg] = c [Vol%] * p_abs /100
Where p = partial pressure, c = gas concentration, p_abs = pressure in breathing circuit,
p_H
O = 21mmHg, partial pressure of water vapor of room temperature gas (23
2
For all other gases the readings are always given as dry values.
Sample Flow Rate: 200 ml/min
Sample Delay Time: All measurements and alarms are subject to a delay of 5 seconds.
Total System Response Time = the sum of the delay time and the parameter specific rise time.
1. After warm up or zero the flow rate may be higher than 200 ml/min for about 30 minutes.
10 to 40C (50 to 104F)
-20 to 65C (-4 to 149F)
5 to 90% RH max @ 40C (104F).
non-condensing
5 to 95% RH max @ 65C (149F).
non-condensing
-305 to 2900m (-1,000 to 9,515ft)
-305 to 5000m (-1,000 to 16,404ft)
1-2 minutes to measure CO
accuracy specifications
the humidity correction can be set to "wet" or "dry".
2
O)/100
2
1
20
1 Introduction
, less than 6 minutes for full
2
o
C, 100% rh).
9