4. To make sure that the calibration is correct, select the
applicable mV input or output task (Table 2) and apply
the necessary values:
• mV input: -10, -5, 0 (short circuit)
Then mV: 5, 10, 20, 50, 100
• mV output: -10, -5, 0, 5, 10, 20, 50, 100
5. Make sure that the error is in the specified limits
(Table 10).
Table 10: mV input or output error limits
Input
Calibrator error
or output
(mV)
(mV)
mV
±10
0.0005
±5
0.0004
0
-
20
0.0006
50
0.0008
100
0.0012
Calibration - Procedures (CJ input)
1. Connect the instrument to the calibration equipment
(Figure 1).
2. Let the equipment get to a stable temperature
(minimum: 5 minutes since the last power on).
3. Use the calibration menu (Table 9) to do a one-point
calibration (+FS). The display shows the applicable
instructions to complete the calibration.
4. To make sure that the calibration is correct, select the
applicable T1 input task (Table 2).
5. Make sure that the DPI 821/822 gives a probe
temperature that agrees with the temperature on
the digital thermometer ±0.2°F (0.1°C).
Calibration - Procedures (mA input)
1. DPI 822 only. Connect the instrument to the calibration
equipment (Figure 5).
2. Let the equipment get to a stable temperature
(minimum: 5 minutes since the last power on).
3. Use the calibration menu (Table 9) to do a two-point
calibration (-FS, Zero and +FS). The display shows the
applicable instructions to complete the calibration.
K346 Issue 2
Permitted DPI 821/822
error (mV)
mV
mV
0.00018
0.010
0.010
0.00011
0.009
0.009
0.00005
0.008
0.008
0.00030
0.011
0.011
0.00068
0.016
0.016
0.00130
0.024
0.024
4. To make sure that the calibration is correct, select the
applicable mA input task (Table 2) and apply these
values:
• mA: -55, -40, -24, -18, -12, -6, 0 (short circuit)
Then mA: 0, 6, 12, 18, 24, 40, 55.
5. Make sure that the error is in the specified limits
(Table 11).
Applied
mA
mV
±55
±40
±24
±18
±12
±6
0 (Short circuit)
Calibration - Procedures (IDOS UMM)
Refer to the user manual for the IDOS UMM.
When the calibration is complete, the instrument
automatically sets a new calibration date in the UMM.
Specification data
All accuracy statements are for one year.
Specification - General
Languages
Operating
temperature
Storage
temperature
Humidity
Shock/Vibration
EMC
Safety
Size (L: W: H)
Weight
Power supply
Duration
(Measure)
Duration
(Supply)
Table 11: mA input error limits
Calibrator
Permitted
error
DPI 821/822 error
(mA)
(mA)
0.0022
0.005
0.0018
0.004
0.0014
0.003
0.0004
0.003
0.0003
0.002
0.0002
0.002
-
0.001
English [Default]
14 ... 122°F (-10 ... 50°C)
-4 ... 158°F (-20 ... 70°C)
0 to 90% without condensation
(Def Stan 66-31, 8.6 cat III)
BS EN 61010:2001; Def Stan 66-31, 8.4 cat III
BS EN 61326-1:1998 + A2:2001
Electrical - BS EN 61010:2001; CE Marked
7.1 x 3.3 x 2.0 in
(180 x 85 x 50 mm)
14 oz (400 g)
3 x AA alkaline batteries
T1, mV: ≈ 70 hours
mA: ≈ 25 hours
mA: ≈ 11 hours (24 V Source at 12 mA)
T1, mV: ≈ 70 hours
[EN] English - 11