Specifications

Sensor input capability
Operating humidity
Operating temperature
Storage temperature
Power requirements
Enclosure
Display visibility
Clock
Mounting
Shipping weight
Outputs (AquaTrend with SOM only) Two current analog outputs; range either 4-20 mA or 0-20 mA (user selectable);
Relays (AquaTrend with SOM only)
Communications
Data logging capability
AquaTrend Interface and AquaTrend/SOM
Function
Accessories
Page 6
Specifications
Specifications are subject to change without notice.
Eight
5 to 95% non-condensing
0 to 40 °C (32 to 104 °F)
-20 to 60 °C (-4 to 140 °F)
External power supply 12 VDC nominal ±25%
(9 to 15 VDC range) with peak to peak ripple to be less than 10% of nominal.
NEMA 4X (indoor), IP 66
LCD backlit, visible at 10 feet with single reading; visible at 3 feet with multiple
readings.
Real time
Wall, panel, or pole
®
AquaTrend
Interface or AquaTrend with Signal Output Module (SOM):
4 lb Portable AquaTrend Interface: 3 lb
both outputs isolated to 500 VDC; may be trimmed ±20% if needed (via AquaTrend
Interface).
SPDT 250 VAC, 5A resistive
®
Echelon
fieldbus compatible; utilizes LonTalk
between devices with a 500 meter maximum per segment; distances in excess of
500 meters require a repeater. Up to 3 repeaters can be used for a total network
length of 2000 meters.
1 hour log (1 min. intervals), 30 day log (15 min. intervals)
Capable of Master or Remote operation; the Master AquaTrend Interface is capable
of communicating with 34 Hach devices including:
8 sensors and/or analyzers
8 Signal Output Modules (SOM)
8 Digital Display Modules (DDM) or 8 Portable AquaTrend Interfaces
8 Remote AquaTrend Interfaces
2 Serial I/O Modules (SIO)
Network configuration, system setup, sensor calibration, data logging
Network terminator provided on the Signal Output Module (SOM) board or the
AquaTrend Interface. One required per network.
®
protocol. Maximum of 400 meters
51350-18 specifications.fm