External Network Manager - Hach AquaTrend 51200 Instrument Manual

Table of Contents

Advertisement

Section 3

3.11 External Network Manager

Page 40
External Network Manager
4. Choose the channel (A–P) for which the data is to be printed by moving
the pointer to the appropriate channel and pressing ENTER.
5. Choose the timebase of the data to be printed.
a. Select TIME BASE and press ENTER.
b. Move the pointer to the appropriate time base and press ENTER.
6. A pop-up display will show the instrument, timebase and type of values to
be printed. Press ENTER to continue.
If an external network manager is used, internal network management of the
AquaTrend must be disabled.
Access the Network Manager function and disable AquaTrend network
management as follows:
1. Perform the AquaTrend cold start procedure. Refer to
page 72
for instructions.
2. Select MASTER and press the ENTER key.
3. Select NETWORK MANAGER OFF and press the ENTER key.
An external manager may now access the Master AquaTrend Interface.
section 6.2.6 on
51350-18 Initial Aquatrend Setup.fm