3.12 Restore Net Connections

51350-18 Initial Aquatrend Setup.fm
The Restore Net Connections Menu option is used to restore all device
connections previously on the network to a Master AquaTrend Interface
functioning as the Network Manager. The data log information will be lost; that
information should be recorded or down loaded to a printer or personal
computer via an SIO (with appropriate software, such as Windows 3.1
Terminal or Windows 95 Hyperterminal) before the Restore Net Connections
procedure is initiated.
Restore Net Connections
Begin the
1. Perform a cold start on the AquaTrend Interface to remove any residual
information which may affect the recovery process. Refer to
on page 72
for instructions. If a replacement AquaTrend Interface is
being installed, physically attach the AquaTrend Interface to power and to
the network.
2. From the initial AquaTrend Interface screens, select MASTER and
press ENTER.
3. Select NETWORK MANAGER and press ENTER.
4. Press the MENU key to return to the Main Menu.
5. Use the arrow keys to move the pointer to Network Menu.
6. Select RESTORE NET CONNECTIONS and press the enter key. A list of
devices found will be shown. If the list of devices appears correct, press
the ENTER key.
Note: If the list of devices is incomplete, press the EXIT key. Make sure all devices are
on-line and powered and repeat Step 6. If all expected devices are still not found,
press ENTER and continue this procedure. Missing devices may be added later
using the Add Device procedure in
7. A progress indicator screen will appear as the network is restored. Press
the ENTER key to complete the procedure.
8. Select REVIEW NET CONNECTIONS and press the ENTER key.
procedure as follows:
section 3.4.2 on page
26.
Restore Net Connections
Section 3
section 6.2.6