Table 7: General Specifications
DEVICE
Size (H x W x D)
Weight
Power
Device Classification
Design Standards
ENVIRONMENT
Operating Temperature
Storage Temperature
Humidity:
Vibration
Shock
Altitude
Particle and Water Ingres
DEFIBRILLATOR
Waveform
Defibrillator Charge Hold
Time
Energy Selection
Patient Safety
Charge Time
Electrodes
Built in Defibrillator Self Test
5.25"x 9.50" x 11.50"; 13.3 cm x 24. 1 cm x 29.2 cm
6.7 lbs.; 3.1 kg
User Replaceable Batteries. 10 -Type 123A Photo Flash lithium
manganese dioxide batteries.
Class II and internally powered per EN60601-1
Meets applicable requirements of UL 2601, AAMI DF-39, IEC
601-2-4, EN60601-1, IEC60601-1-2
PS Model: 32 °F to 122 °F; 0 ο to 50 ο C
PA Model: 50 ο to 104 ο F; 10 ο to 40 ο C
PS Model: -22 ο to 158 ο F; -30
PA Model: 32 ο to 122 ο F; 0 ο to 50 ο C
10 to 95% relative humidity, non-condensing
MIL Std. 810F, Min Helicopter Test
PS Model: IEC 68-2-27; 100G
PA Model: IEC 68-2-27; 50G
PS Model: -300 to 15,000 ft.; -91m to 4573m
PA Model: -300 to 7,500 ft.; -91m to 2287m
IP-55
Rectilinear Biphasic
30 seconds
Automatic pre-programmed selection (120J, 150J, 200J)
All patient connections are electrically isolated.
Less than 10 seconds with new batteries.
ZOLL stat padz II or CPR-D padz.
Included
ZOLL AED PLUS