Adam Equipment WBZ Series Service Manual page 12

Table of Contents

Advertisement

F2
100g
F5
Point
F6
init
F7
rE
F8
SCSivE
F9
Ad
F10
round
F11
ZEro
F12
Lvd
F13
CAL
F14
ACLr
F15
CAL
Other parameters must be changed using the service parameters as
described in section 8.2
After calibration or changing of any parameters the value of the counters is
required to be visible on the scale marked in a permanent manner. Adam
Equipment supplies a label for this purpose, to be placed near the verification
marks on the side of the scale.
Any time these values change the scale is required to be retested by the
appropriate authorities in the users location.
© Adam Equipment Company 2008
Enable price per 100g or per kg option
Unit Price/Price decimal point
ZEro
Initial Zero Range
ZEro
Re-Zero range
tArE
Successive Tare Enable
Count
Display the A/D counts
Rounding off the Total Price
Select the zero mode
mod
Low voltage detection
batt
Set calibration of battery voltage measurements
Not used, Auto Clear is always enabled
Cnt
Enable or disable the calibration counters
11